Ocv.StopAt()

Ocv.StopAt()

Type: INSTANCE FUNCTION

Purpose: Stop the curve if either the rate of change in E is too small or too great.

The first case is used to terminate a curve when the sample has stabilized. The second test is useful in detecting an unstable sample.

NOTE: There is a default 10-point delay on any stop test because of the inherent Bessel filter used in the OCV curve.

1. Stable d(Eabs)/dt < dE/dt min

2. Runaway d(Eabs)/dt > dE/dt max

Usage: Ocv.StopAt(Stable, Runaway)

  Stable

REAL

or

Enable the Stable test and set value .
  NIL Disable the Stable test.
  Runaway

REAL

or

Enable the Runaway test and set value.
 

NIL

Disable the Runaway test.

Related Topics Link IconRelated Topics