Cyclic Polarization Experimental Sequence
Run the Cyclic Polarization standard technique by selecting Experiment > DC Corrosion > Cyclic Polarization from the Framework™ menu bar. This initiates the following sequence of events:
- Framework creates a Runner window, and the Cyclic Polarization.exp script is run in this window.
- The script creates the Setup dialog box and accepts changes in the experimental parameters.
- The script now obtains the use of the potentiostat specified during Setup and opens the data file using the Output name. If the potentiostat is in use or the file cannot be opened, the script returns you to the Setup dialog box.
- The file header information is written to the data file. This header information includes:
- Tags identifying possible analyses
- The current time and date
- A list of the Setup parameters
This information is written to the file prior to data-acquisition. If the experiment is aborted, the output file contains only this information.
- The script conditions the electrode if Conditioning was specified in the Setup. Conditioning is done by applying a fixed potential for a defined time. It establishes a known surface state on the corrosion specimen. A plot of current versus time is displayed during the conditioning.
- The cell is turned off and the specimen’s Eoc is measured.
- If an Init. Delay is turned on in the Setup, this step lasts for the time specified as the Init. Delay time, or until the potential stabilizes. If no Init. Delay is specified, this step only lasts 10 s. A plot of potential versus time is always displayed. The last measured potential is recorded as Eoc.
- The scan starts and data are acquired. The potential of the sample is swept from the Initial E to the Apex E, where the scan is reversed. The reverse scan is from the Apex E to the Final E. The initial scan can be reversed prior to the Apex E, if the current exceeds the Apex I parameter. The sweep is actually a staircase ramp. The sample is potentiostated at the Initial E, a delay of one Sample Period occurs, and a reading of the current is taken. The potential is then stepped a few mV, a delay of one Sample Period occurs, and the next current reading is taken. If the current exceeds the Apex I or the potential exceeds the Apex E, the reverse scan is started. The reverse scan is very similar to the forward scan, except that the sign of the steps is changed. You can select different Scan Rates for the two scans. The reverse scan continues until the potential reaches the Final E, when the scan is halted. At each point the current range is automatically switched to the optimal range for the measured cell current. If IR-Compensation has been selected, each potential is corrected for the measured IR-drop of the preceding point. A plot of log(I) vs. E is displayed during the scan.
- The data are written to the output file and the script cleans up and halts.
- When the scan is over, the cell is turned off. The acquired data are written to the output file.
- The script waits for you to select F2-Skip.When you do, the script closes everything that is open, including the Runner window.
Comments are closed.