Signal.SetAcquisitionControl()

<< Click to Display Table of Contents >>

Navigation:  Explain™ Control Language  > Library Routines > Input and Output functions >

Signal.SetAcquisitionControl()

Description

Allow user to specify specific data-acquisition parameters.

 

Gamry tip bulb

This instance function is available for all signals.

 

Type

INSTANCE FUNCTION

 

Usage

Signal.SetAcquisitionControl (AcqMode, AcqBasePeriod, SampleRate, DutyCycle)

 

AcqMode

INDEX

Type of acquisition mode:

0 = Fast Mode

1 = DSP Mode

NIL

Use the default AcqMode. This mode is set based on the output point sample period:

Below 100.0 µs the system uses Fast Mode.

At or above 100.0 µs, the system uses DSP Mode.

AcqBasePeriod

REAL

The time between each sample point. This number should not be less than the default of 16.666 µs.

NIL

Use the default AcqBasePeriod:

For ACQ_MODE_FAST, this is not used.

For ACQ_MODE_DSP, the default is 16.666 µs.

SampleRate

REAL

The output point rate, in seconds. Output points are the points displayed to the user and saved to the curve. They are made up of one or more sample points.

NIL

Use the default sample rate specified in the SIGNAL.New call.

DutyCycle

REAL

A number between from 0.0 to 1.0 specifying the DutyCycle that is used in sampling:

0.0 always gives one sample.

1.0 samples as quickly as possible.

INDEX

0 = One sample only (DutyCycle of 0.0).

1 = Use the default DutyCycle of 0.20.

2 = Use a DutyCycle of 1.0.

NIL

Use the default DutyCycle of 0.20.

 