Pwr.SetStopADAMin()
Type: INSTANCE FUNCTION
Purpose: Set a Stop At based on the stability of the Aux channel measurement.
The ADAMin test checks that the abs(dA/dt) < Limit. Once met, the data acquisition skips to the next signal section, or stops if there is no subsequent signal section.
Usage:
Pwr.SetStopADAMin(Limit)
Limit
- REAL
- Limit value entered in Volts/second.