Pwr.SetStopAtDelayAAMax()

Type: INSTANCE FUNCTION

Purpose: Used to avoid premature stopping of a curve because of noise creating a false reading for a StopAt test.

Requires StopAt criterion be met for a number of data points (N) before the curve stops. The parameter N can be individually set for each StopAt test. This function sets the delay for the AAMax stop test. The criterion must be met for N points consecutively before the criterion will be accepted. One point missing the criterion limit resets the N point counter.

Usage:

Pwr.SetStopAtDelayAAMax(Delay)

Delay

  • INDEX
  • Delay value entered in number of points.

Related Topics Link IconRelated Topics