Galvcor.Thresh()
Type: INSTANCE FUNCTION
Purpose: Set up threshold conditions which must be met before Galvcor.StopAt() tests are performed.
When any of the threshold tests is met, the StopAt criteria are evaluated for the rest of the scan. There are six threshold tests that you can perform:
1. IMin
I < I_MinThresh
2. IMax
I > I_MaxThresh
3. EMin
E < E_MinThresh
4. EMax
E > E_MaxThresh
5. Tmin
T < T_MinThresh
6. Tmax
T > T_MinThresh
Threshold tests allow you to tune the StopAt tests more carefully. For example, you can avoid a premature experiment termination caused by an initial transient.
In general you only enable one of the threshold tests. The others are turned off by a NIL argument in the function call.
Note: Any threshold test starts all StopAt tests. Also, when StopAt testing is enabled, it is never disabled, even if the Threshold test that enabled it is no longer valid.
Usage:
Galvcor.Thresh(IMin, IMax, EMin, EMax, TMin, TMax) Imin or REAL
Enable StopAt if I < IMin
NIL Disable this test. Imax or REAL Enable StopAt if I > IMax NIL Disable this test. EMin or REAL Enable StopAt if E < EMin NIL Disable this test. EMax or REAL Enable StopAt if E > EMax NIL Disable this test. TMin or REAL Enable StopAt if T < TMin NIL Disable this test. TMax or REAL Enable StopAt if T > TMax NIL Disable this test.
Comments are closed.