Pstat.SetAchSelect()

Pstat.SetAchSelect()

Type: INSTANCE FUNCTION

Purpose: Pstat.SetAchSelect specifies the input to be measured by the Auxiliary channel on the control board.

The default setting is the BNC input on the control board

Usage: AchSelectSet = Pstat.SetAchSelect (Input)

 

       
  Input INDEX

Input to select

0 = BNC connector (Default)

1 = Pstat (Not used currently)

2 = Control Amplifier on potentiostat

3 = Thermocouple

4 = Ground

  AchSelectSet INDEX The new input selected

This is valid for PC5 instruments.