Galvanostatic Mode Electrochemical Noise

<< Click to Display Table of Contents >>

Navigation:  Experiments > H - Electrochemical Noise > Techniques >

Galvanostatic Mode Electrochemical Noise

Description

An Electrochemical Noise test in galvanostatic mode measures potential noise at a fixed current. A specimen is immersed in a solution and a constant current is applied while measuring its potential (versus a reference electrode) in a three-electrode experiment.

 

Galvanostatic Mode ECN is useful when you are modeling the surface fluctuations that affect the charge-transfer processes of your system at a steady-state current. ZRA Mode ECN cannot test this type of system.