Charge

<< Click to Display Table of Contents >>

Navigation:  Experiments > J - Electrochemical Energy > Techniques >

Charge

Description

This script is used to charge an electrochemical cell to a set voltage using constant current. For typical energy-storage devices, this involves a reverse process where electrons flow out of the (discharge) cathode and into the (discharge) anode. It allows an optional voltage finish during which the voltage is maintained for a user-specified amount of time or until the current-level falls below a user-specified value.

 

This is a half-cell, full-cell, stack, or dual-electrometer experiment. Dual-electrometer mode is only available with the Interface 5000.

 

Related Topics