Potentiostat Capabilities
Potentiostats
Gamry Instruments offers two lines of potentiostats, the Reference™ line and the Interface™ line. Our Reference line of potentiostats are our high-performance potentiostats designed to give you the absolutely best measurement from low currents to high currents and anything in between. Our Interface instruments are more general-purpose, and combine both value and performance in one package. Though they do not have the same specifications as the Reference line, they are great for a large number of electrochemical experiments.
Each potentiostat is able to perform different sets of experiments. Below is a table summarizing our potentiostats’ capabilities.
Error Messages and Limitations
Interface 5000P
The Interface 5000 has a subtype, Interface 5000P, which may generate error messages under certain run-time conditions.
Error Message CntrlMode=0 | If you see a CntrlMode=0 error, this means that you tried to run a potentiostatic experiment on an Interface 5000P, which is not able to run such scripts. The Interface 5000P may only run galvanostatic experiments. |
Error Message This instrument is not authorized for Freq= | The Interface 5000P is able to perform EIS at frequencies 20 kHz or lower. Higher frequencies generate this error message. Decrease the upper frequency to continue |
Interface 1000T
The Interface 1000 has a subtype, Interface 1000T, which may generate error messages under certain run-time conditions.
Error Message This instrument is not authorized for Freq= | The Interface 1000T is able to perform EIS at frequencies 20 kHz or lower. Higher frequencies generate this error message. Decrease the upper frequency to continue. |
Error Message This instrument is not authorized for IERange= | The Interface 1000T is able to operate at currents of 100 mA or lower. Higher currents generate this error message. Decrease the current to continue. |
Error Message This instrument is not authorized for ScanRate= | The Interface 1000T is able to scan at rates of 1 V/s or lower. Higher scan rates generate this error message. Decrease the scan rate to continue. |
Error Message This instrument is not authorized for SampleTime= | The Interface 1000T is able to record data with a sample time of 1 ms or higher. Shorter sample times generate this error message. Increase the sample time to continue. |
Comments are closed.