Cyclic Charge Discharge - Setup Parameters (page 3)

<< Click to Display Table of Contents >>

Navigation:  »No topics above this level«

Cyclic Charge Discharge - Setup Parameters (page 3)

Description

Second setup page of the Cyclic Charge Discharge experiment. Press OK to continue or Cancel to go back to the second setup page.

 

7 - PWR CCD - page 3

 

Parameter

Description

Units

Save Interval

Set the interval for which Raw Data files are saved:

All: Save all data files.

Skip N: Save every (N + 1) cycle, skipping over Skip Number N data points. For example, N = 4 saves every fifth data point.

After Delta Q: Save data after a certain percentage change in charge Q specified in Delta Q Increase and Delta Q Decrease.

 

Skip Number

Specify the frequency at which raw data files are saved. If this number is zero, files are saved for every charge-discharge cycle. If the Skip # is set to a value greater than zero, n cycles are skipped for each cycle that is saved.

 

For example, if Skip # is set to 4, the 1st cycle is saved, the next four cycles (2, 3, 4, and 5) are skipped, cycle 6 is saved, cycles 7, 8, 9, and 10 are skipped, cycle 11 is saved, and so on.

 

Delta Q Increase

The percentage change in charge Q which must occur before the system starts recording raw data points, if you have selected Yes to Save Raw Data.

percent (%)

Delta Q Decrease

Initial Freq.

The starting frequency of the frequency sweep during data acquisition.

 

Gamry tip bulb

EIS scans are usually run with the Initial Freq. larger than the Final Freq. parameter. Refer to the potentiostat's Operator's Manual for detailed information on the applicable frequency range.

hertz (Hz)

Final Freq.

The final frequency of the frequency sweep during data acquisition.

 

Gamry tip bulb

The frequency sweep may not stop exactly at the final frequency. It is mathematically impossible to control both Points/decade and Final Freq. parameters exactly for all scan ranges. The EIS software chooses to control the Points/decade parameter exactly.

hertz (Hz)

Points/decade

The data density of the measured impedance spectrum. The data are spaced logarithmically and the number of data points in each frequency decade equals Points/decade. As a consequence, the frequency sweep may not stop exactly at the final frequency. It is guaranteed to do so only when the scan range contains an integer number of decades, such as 5 kHz to 0.05 Hz (five decades). You can use Initial Freq., Final Freq., and Points/decade to calculate the total number of data points in the spectrum.

 

 

DC Current

The constant current applied to the cell throughout the frequency sweep. The AC Current is summed with the DC Current. In most cases, the DC Current should remain at its default value of zero. The discussion of the AC Current describes some limitations on the DC Current value.

amperes (A)

AC Current

The amplitude of the AC signal applied to the cell. Multiply the entered root-mean-square (rms) value by √2 (or ~1.414) to convert into a peak value.

 

To convert the entered root-mean-square (rms) value into a peak-to-peak value, multiply by 2√2 (or ~2.83).

 

amperes (A) rms

Estimated Z

A user-entered estimate of the cell's impedance at the Initial Freq. parameter. It is used to limit the number of trials required before acquiring the first data point in an impedance spectrum. It is generally sufficient if Estimated Z is within a factor of five of the cell's impedance.

 

ohm

Optimize for

Select the sampling method for the experiment:

Fast is the appropriate selection when the cell's stability is poor and a spectrum must be measured rapidly, or the system's impedance is low and well defined.

Normal is the appropriate selection when the cell's impedance is high or the electrochemical system is noisy.

The best data can be taken with Low Noise, but the time required to record an EIS spectrum can be quite long.

 

 

Start on Cycle

Set the cycle number of the first EIS measurement.

 

Save Interval

Set the interval for which Raw Data files are saved:

All: Save all data files.

Skip N: Save every (N + 1) cycle, skipping over Skip Number N data points. For example, N = 4 saves every fifth data point.

After Delta Q: Save data after a certain percentage change in charge Q specified in Delta Q Increase and Delta Q Decrease.

 

Skip Number

Specify the frequency at which raw data files are saved. If this number is zero, files are saved for every charge-discharge cycle. If the Skip # is set to a value greater than zero, n cycles are skipped for each cycle that is saved.

 

For example, if Skip # is set to 4, the 1st cycle is saved, the next four cycles (2, 3, 4, and 5) are skipped, cycle 6 is saved, cycles 7, 8, 9, and 10 are skipped, cycle 11 is saved, and so on.

 

Delta Q Increase

The percentage change in charge Q which must occur before the system starts recording raw data points, if you have selected Yes to Save Raw Data.

percent (%)

Delta Q Decrease

THD

Enable Total Harmonic Distortion (THD) during the EIS experiment to obtain additional information about the system’s harmonics.

 

Drift Correction

Select On to enable Drift Correction. Both the original and the drift-corrected impedance values are separately calculated and recorded. The drift corrected impedance plot will be active by default but can be changed from the drop-down menu above the chart.

 

Drift Correction fits current and voltage data to a sine wave using a linear drift term, followed by non-linear least squares regression. Drift data are then subtracted from the current and voltage values and the corrected impedance Z is calculated via Fourier analysis.

 

Gamry tip bulb

Drift Correction is always disabled if an overload is detected, the frequency is above 10 Hz, or cycle decimation is occurring. Drift most commonly occurs at lower frequencies, which is why limitations are applied to this frequency range.

 

 

Related Topics